Reliability and Validity of Test of BTE Primus Grip Attachment

OUYANG Ya-tao, XUYan-wen   

  1. Guangzhou Work Injury Rehabilitation Hospital, Guangzhou 510970, Guangdong, China
  • Received:2007-05-21 Published:2007-08-01 Online:2007-08-01

摘要: 目的测试BTE Primus多功能康复系统握力器的重测信度和效标效度。方法受试者为30名成年健康男性,平均年龄(22.23±1.59)岁。测试受试者两次使用BTE握力器和Jamar握力器的力量,两次测试间隔时间为7d。以等级相关系数(ICC)作为重测信度指标;相关性系数r作为效标效度指标。结果BTE Primus握力器的测试结果与Jamar握力器的结果高度相关(r=0.842,P<0.01),且重复测试信度较高(ICC=0.957,95%CI=0.908~0.980)。结论BTE握力器可在临床手功能评定中使用。

关键词: 信度, 效度, 握力, BTE Primus系统

Abstract: Objective To test the test-retest reliability for the BTE Primus grip attachment.Methods30 healthy adults were recruited to test the grip strength of BTE Primus grip attachment and Jamar grip tool in one day, the 2nd retest was conducted 7 days late.ResultsThe BTE Primus grip tool was found to be reliable (ICC = 0.957, 95%CI = 0.908~0.980) and valid ( r=0.842, P<0.01).ConclusionBTE Primus grip attachment can be applied when evaluating hand function in clinic.

Key words: reliability, validity, grip, BTE Primus system