Reliability of Ultrasound Tissue Characterization for Quantifing Integrity of Achilles Tendon and Patellar Tendon in People without Tendinopathy

LUO Ping1, LIN Hong-sheng2, FANG Jian-hui3, ZHANG Wen3, CAI Chao-chen4   

  1. 1. Department of Sports Health Care, Guangdong Vocational Institute of Sport, Guangzhou, Guangdong 510663, China;
    2. Department of Clinic, Special Operation Academy, Guangzhou, Guangdong 510500, China;
    3. Guangdong Sports Hospital, Guangzhou, Guangdong 510100, China;
    4. Department of Rehabilitation Sciences, The Hong Kong Polytechnic University, Hong Kong 999077, China
  • Received:2017-05-10 Revised:2017-08-01 Published:2017-10-25 Online:2017-11-03
  • Contact: LUO Ping. E-mail: cptlp@163.com

Abstract: Objective To investigate the reliability of ultrasound tissue characterization (UTC) for quantifing the integrity of achilles tendon and patellar tendon in subjects without tendinopathy.Methods From August, 2016 to April, 2017, testers A and B quantified integrity of achilles tendon and patellar tendon of dominant legs from 43 subjects without tendinopathy by UTC in random order respectively. Tester B quantified integrity of patellar tendon of dominant legs from other 22 subjects without tendinopathy by the same way, and repeated the measurements the next day. Interclass correlation coefficient (ICC) and the minimum detectable different (MDD) were calculated.Results The ICC of inter-rater reliability of achilles tendon assessment was 0.492 to 0.735, MDD was 0.3% to 3.0%. The ICC of inter-rater reliability of patellar tendon assessment was 0.383 to 0.678, MDD was 0.3% to 4.4%. The ICC of intra-rater reliability of patellar tendon assessment was 0.525~0.760, MDD was 0.6% to 5.3%.Conclusion The test-retest reliability of UTC for the quantification of integrity structure of achilles tendon and patellar tendon is satisfactory with little error, the inter-rater reliability need to be improved.which is a accurate and feasible tool to quantify tendon by the same tester.

Key words: ultrasound tissue characterization, achilles tendon, patellar tendon, reliability